Conductivity sensor type M8836S-10
  • Field of application: Water- and waste water conditioning
  • internal Pt-100 temperature sensor
  • cell constant K=10.0
  • cell material: PVDF
  • measurement range: 0.05mS to 200mS
Data Sheet
Certificate
Product Details

The conductivity cell M8836S10 is used for water and waste water measurements in batch and continuous applications. They are used e.g. in chromatography, water purification fermentation and general chemical processes.

The accuracy is 3% in the temperature range of -30...100°C.

The cell constant K is 10.0 for the measuring range of 0.05...200mS. It is unsuitable for waste water with conductive dirty particles.

The insulating parts are made in PVDF and leak proofing is guaranteed by chemically resistant O-rings seals.

A build in Pt-100 (RTD) temperature sensor enables automatic temperature compensation.